The new Leach electronics were installed at the Southern 2MASS Facility on 990226 UT. Two dark scans were taken and the results were analysized by Mike Skrutskie and were also run through the normal DARKS subsystem in 2MAPPS. Mike reported that the darks appeared similar to those produced when the electronics were tested in the lab. The only exception was some minor amplifier glow in the corners of the arrays. He stated that the amplitude of this glow should be well below the level to impact the noise in the darks. The images showing the difference between the darks taken with the Leach system and recent souther darks taken with the GATIR system show the glow in the corners (see below). The amplitude of the signal in the corners relative to the middle of the arrays is ~50-60 DN in J, ~40 DN in H and ~30 DN in Ks.
The dark frame statistics reported by DARKS is as follows:
Band/ReadType: J/1 H/1 K/1 J/2 H/2 K/2 (evening) Avg Dark Sigma: 4.57 4.31 4.24 2.20 2.24 1.94 St Dev of Above: 0.85 1.38 1.28 0.58 2.82 1.41 No. Masked Pix: 0 0 0 6 138 24
The images below compare the new dark frames (on the left) with
recent southern canonical darks (on the right), taken from the
night of 990217s UT.
J-band (left-990226s right-990217s)

H-band (left-990226s right-990217s)

Ks-band (left-990226s right-990217s)

These three images show the differences between the
990226s and 990217s dark frames in each, band. All are shown with
a fixed display range between 0 and 70 DN.
J-band

H-band

Ks-band

The electronics were installed on the telescope for night of 990227 UT. A full night of observations were made under nominally photometric conditions. Two calibration scans, scans 045 and 046 covering the S312-T cal field, and one survey scan were ftp'd to IPAC and processed using 2MAPPS v2.1. The pipeline output from these runs was evaluated and summarized below.
There is no evidence of the "anti-ghosts" seen near bright stars when the Leach electronics were first installed on the northern telescope.
From Mike Skrutskie:
I was able to generate the equivalent of flat frames locally and they look perfect. The only problem that could arise now would come from the statistics of source counts and uncertainties, but that is unlikely at this point. ...
From Gene Kopan:
Southern telescope with Leach electronics installed.
Looked at elvis:/k4/OPS/990227s/s046-s047, r1 noise & r2 flattened
frames look ok, these scans are fairly high density ( ~5000 extrations- cal
scans) so subltle effects in the r2 may not be obvious, & noise is pumped
up by confusion noise.
Sawtooth pattern in r1 frame medians, peak-to-peak:
j 1.0 dn
h 0.5 dn
k 1.5 dn
Solos about normal for this density, frexas residuals look ok. No
obvious problems. It would be good to check the r2 noise on a low density
scan.
S045 S046
Mcat Minst ZP Minst ZP
-----------------------------------------------------------------
J 11.949 11.988 +/- .021 -0.039 12.106 +/- .029 -0.157
H 11.669 11.642 +/- .031 +0.027 11.632 +/- .021 +0.037
Ks 11.609 11.580 +/- .029 +0.029 11.580 +/- .027 +0.029
The zero point offsets for this cal scans are slightly higher, but generally consistent with recent values measured at CTIO. The zero point offsets from the night of 990217s are: -0.187 mag, -0.072 mag, and -0.057 mag at J, H and Ks, respectively.
The H and Ks instrumental magnitudes repeat extremely well, but the J magnitudes differ by almost 10% between the two scans. The figure below shows the magnitude difference between all stars common to the two scans, plotted versus their average magnitude. The heavy lines show the trimmed-average magnitude difference within 0.5 magnitude bins. The vertical error bars show the RMS magnitude difference in each bin. There is no statistically significant bias in the instrumental photometry between the two scans. Note that no effort has been made to remove confused objects, and this is a moderately dense field. Note that confusion noise has moved the SNR=10 level close to the Level 1 Req. brightness.
Magnitude differences plotted versus mean magnitude for all
stars in common between S045 and S046

Photometric Repeatability from 8 scans covering the S312-T Cal Field

The band-band photometric performance appears nominal, as shown by the color-color plots for scans S040, S045 and S046, shown below. These are made using instrumental magnitudes, only, so there is some slight offset in J-H due to the relative zero-point offsets. (flagged artifacts have been removed, but no attempt has been made to remove confused sources). In all of these plots, objects with SNR(Ks) > 20 are shown in blue.
Instrumental color-color plot for S040

Instrumental color-color plot for S045

Instrumental color-color plot for S046
